Traceability excerpt: audit trail and electronic signatures
Laboratory results system, a configured product (GAMP 5 Category 4). Each user requirement traces to the risk that sets how deeply it is tested, the tests that cover it, and the evidence that they passed.
- URS-014Risk priority High
The system records who changed a result, when, the old and new value, and why, in an audit trail that users cannot edit or delete.
- Risk
- RA-07 A change with no record could hide a data integrity failure in a released result.
- Tested by
- OQ-021 Change a result as an analyst. The audit trail shows the user, time, old and new value, and reason.Pass
- OQ-022 Try to edit and delete an audit trail entry as an administrator. The system refuses both.Pass
- Evidence
- Signed OQ execution record, with the audit trail export and screenshots attached.
- URS-022Risk priority Medium
Releasing a result needs an electronic signature that shows the signer's name, the date and time, and the meaning of the signature.
- Risk
- RA-11 A release that cannot be traced to a person. The signature function is standard and not changed by configuration.
- Tested by
- OQ-030 Release a result with an electronic signature. The screen and the printout both show name, date and time, and meaning.Pass
- Evidence
- Signed OQ execution record, plus the supplier's test report for the signature function, reviewed and referenced.